MODEL 7550In MotionConveyor ScaleINSTALLATION MANUAL

Pagina 2

The WI-127's front panel consist of 24 keys and 14 annunciators.The WI-127's keys are divided into four primary groups:Standard Scale Keys –

Pagina 3 - TABLE OF CONTENTS

Operation of these keys changes with thesoftware installed in the WI-127 (standard,in-motion).Annunciators8Function Keys – These oval keys along the

Pagina 4 - 7550 Description

Power Requirements:115 Volts AC, +10% to -15% @ 0.3Amp maximumExcitation:10 Volts DCSupports up to twelve 350-ohm weight sensorsOperational Keys:Five

Pagina 5 - Installing the Conveyor Scale

10Standard Inputs:Seven configurable logic level inputs for functions such as PB tare, print,zero, units, select, gross, net, entrance and exit photoe

Pagina 6 - Conveyor Scale

Error Messages11FlashingFlashingFlashingFlashingFlashing * * * * *SealedSealedSealedSealedSealedCan'tCan'tCan'tCan'

Pagina 7 - Weigh Bar J-Box

Configuring, calibrating and testing of the WI-127 is done using a menustructure which you move through using a directional key menu structure onthe f

Pagina 8

Warning:If you do not press SELECT SELECT SELECT SELECT SELECT with "SAVE""SAVE""SAVE""SAVE""SAVE&quo

Pagina 9 - Wiring Diagram

To calibrate the indicator, you must enter the calibration menu. See note atleft before following these instructions:1. From weight display mode, key

Pagina 10 - WI-127 Indicator Introduction

7. Place the appropriate weight calibration value on the scale and press…"BUSY" is displayed momentarily while the unit obtains a stableval

Pagina 11 - Annunciators

12. Key in the count value for the A XXXXX calibration point.…"VALUE" is displayed.13. Press …Value is accepted and """&q

Pagina 12 - Indicator Specifications

05/30/01 PN 7424-16952A Printed in USARisk of electrical shock. Do not remove cover.No user serviceable parts inside. Refer servicing to qualified

Pagina 13

Deleting Calibration PointsTo exit back to normal weighing mode, pressthe SELECT SELECT SELECT SELECT SELECT key to save changes.5. To add more calibr

Pagina 14 - Error Messages

Follow the instructions in this section to disassemble the WI-127.1. Unplug the WI-127 from the power source.2. Remove the back of the WI-127 by remov

Pagina 15 - Service Menu Structure

Viewing and Setting Time and Date1. From display mode, press MENU twice…"HOUR" is displayed.2. Press ENTER…The current time is displayed.In

Pagina 16 - Calibrating the WI-127

This section of the Installation Manual covers the optional in-motionsoftware for Model 7550. This software has the following additionsmade to the ser

Pagina 17 - Calibration Menu

Getting Weight – There are two (2) ways to get weight from the NCI 7550In-Motion Conveyor Scale. "Standard" or poll method requires a comman

Pagina 18 - Count Calibration

22Dynamic Weighment Serial Output – The gross and net weights in thelayout menus follow the in-motion weighments while in the in-motion mode.Each time

Pagina 19 - Adding Calibration Points

7550 OperationThe in-motion system will only weigh and transmit data properly if there isonly one package on the scale at a time with packages identic

Pagina 23 - 7550 In-Motion Software

TABLE OF CONTENTS7550 Description ... 1Installing the Conveyor Sc

Pagina 27 - SERVICE MENU

1000 Armstrong DriveFairmont, MN 56031 USATelephone: 507-238-4461Facsimile: 507-238-4195E-Mail: [email protected]

Pagina 28

The Weigh-Tronix conveyor scale is a freestanding, self-contained andself-powered, flat-belt weighing system that can be incorporated intonew or exist

Pagina 29

The scale component of the conveyor scale is suspended by means ofchains from four Weigh Bars.® The Weigh Bar outputs feed into a junctionbox which i

Pagina 30

the front edge blocks it. See Figure 2 for photocell module proper program-ming information. Typically, if they are not programmed properly, thesystem

Pagina 31

5. Material breaks beam of exit photocell…Weight is displayed and transmitted to a peripheral device.This scale uses a junction box (J-Box) where the

Pagina 32

If Number of Turns is positive value, turn potentiometersclockwise. If Number of Turns is a negative value, turn potentio-meters counterclockwise.6.

Pagina 33

WI-127 In-Motion WiringWhen installing the WI-127, the power socketoutlet must be nearby and easily accessible.Model 7550 SystemWiring DiagramThe wiri