67XXU
USB/Serial Scale Series
Service Manual
SOFTWARE VERSION 1.20
AWT35-501145
Issue AE
Vedere la pagina 0
1 2 3 4 5 6 ... 84 85

Sommario

Pagina 1 - Service Manual

67XXUUSB/Serial Scale SeriesService ManualSOFTWARE VERSION 1.20AWT35-501145 Issue AE

Pagina 2

10 67XXU USB/Serial Scale Service ManualThe RS-232 port connector is a 9-pin female.Figure 2.3 RS-232 Connector (Com 2)2.2 Column and Remote Display

Pagina 3 - Table of Contents

67XXU USB/Serial Scale Service Manual 11 3. Run the display cable down the inside of the column using the hole on the top side of the column and pul

Pagina 4

12 67XXU USB/Serial Scale Service Manual5. Secure the cable to the column with the cable hold down strap that was remove in step 1.Figure 2.8 Cable

Pagina 5 - 1.1 About this Manual

67XXU USB/Serial Scale Service Manual 13 7. Install the two screws included with the kit into the screw holes leaving them somewhat loose.Figure 2.1

Pagina 6 - 1.5 Cleaning the Indicator

14 67XXU USB/Serial Scale Service Manual10. Turn the scale upright.Figure 2.13 Column and Remote Display Installed2.3 Mounting the Display to the Sc

Pagina 7 - 1.6 CE Certification

67XXU USB/Serial Scale Service Manual 15 2. Clip the display to the bracket.Figure 2.15 Display Attached to the Bracket3. Pull the cable all of the

Pagina 8 - 2 Installation

16 67XXU USB/Serial Scale Service Manual6. Line up holes of the bracket with the holes on the scale base. Fasten the bracket down with the four provi

Pagina 9 - 2.1 Connections

67XXU USB/Serial Scale Service Manual 17 2.4 Desktop and Wall Display MountingThe indicator is connected to the front of the 67XXU. It can be remove

Pagina 10

18 67XXU USB/Serial Scale Service Manual2.5 MainboardFigure 2.23 shows the main board connectors for the loadcell, RS-232 port, USB port, power suppl

Pagina 11

67XXU USB/Serial Scale Service Manual 19 Table 2.3 USB Connector (USB) Com 1Table 2.4 RS-232 Port Connector (SIO) Com 2Table 2.5 CAL Jumper SetPi

Pagina 12

67xx Serial Scale_s_en_500145.bookBrecknell is part of Avery Weigh-Tronix. Avery Weigh-Tronix is a trademark of the Illinois Tool Works group of compa

Pagina 13

20 67XXU USB/Serial Scale Service Manual3 Specifications3.1 Available Models3.2 Outline DimensionsFigure 3.1 67XXU Dimension DrawingFigure 3.2 67XX

Pagina 14

67XXU USB/Serial Scale Service Manual 21 3.3 Mountingl Bracket l Remote (optional cable required)l Column (option kit required)3.4 Power Supplyl AC

Pagina 15

22 67XXU USB/Serial Scale Service Manual3.7 Communication3.8 Analog Circuitryl 24-bit A/D converterl Conversion Speed: 10Hzl Input range: -15mV to +1

Pagina 16

67XXU USB/Serial Scale Service Manual 23 3.12 Other Main Functionsl Programmable zero rangel Programmable pre-set tare weightl Programmable automati

Pagina 17

24 67XXU USB/Serial Scale Service Manual4 ConfigurationTo set up the indicator, you must first enter the appropriate menu mode. The front panel keys

Pagina 18 - 2.5 Mainboard

67XXU USB/Serial Scale Service Manual 25 4.2 KeysThe keyboard consists of four keys, some of which have multiple functions.Table 4.2 Function of th

Pagina 19

26 67XXU USB/Serial Scale Service Manual4.3 Navigating the Weigh Mode4.3.1 Enter the Weigh ModePress the [HOLD] + [UNIT] keys for 3 seconds. Use the

Pagina 20 - 3 Specifications

67XXU USB/Serial Scale Service Manual 27 4.3.7 Output Data (print to a computer or printer)When scale is stable press the [PRINT] key.4.3.8 Display

Pagina 21 - 3.6 Approvals

28 67XXU USB/Serial Scale Service ManualTable 4.4 Use kg as Primary UnitTable 4.5 Use lb as Primary UnitCalibration Division ValueDisplay Division

Pagina 22 - 3.8 Analog Circuitry

67XXU USB/Serial Scale Service Manual 29 4.3.10 Check Weight (compare feature)To make the weight compare function available, set CONFG-FUNC-COMPAR i

Pagina 23 - 3.12 Other Main Functions

67XXU USB/Serial Scale Service Manual 3 Table of ContentsChapter 1 General Information and Warnings ...

Pagina 24 - 4 Configuration

30 67XXU USB/Serial Scale Service Manual4.4.2 Obtain Piece WeightThere are two ways to obtain the piece weight: Input Piece Weight Directly1. Press t

Pagina 25 - 4.2 Keys

67XXU USB/Serial Scale Service Manual 31 8. Press the [TARE] key to confirm the entry. 8a. If the calculated piece weight is less than 0.5d, the ind

Pagina 26 - 4.3 Navigating the Weigh Mode

32 67XXU USB/Serial Scale Service ManualPositive Peak HOLDWhen USER-HOLD-HLD.MD is set to PS.PEK, the hold mode is positive peak hold mode. When scal

Pagina 27 - 4.3.9 Change Weight Unit

67XXU USB/Serial Scale Service Manual 33 The scale will exit HOLD mode according to the setting of USER-HOLD-HLD.TM. If the time of scale variation

Pagina 28

34 67XXU USB/Serial Scale Service Manual4.9 Serial Port 1/2 (COM1/2) Receiving TestBefore testing the receiving function of COM1 or COM2, a cable is

Pagina 29 - 4.4 Navigating the Count Mode

67XXU USB/Serial Scale Service Manual 35 5 Setup Mode The setup menu consists of five different sub-menus. Within each sub-menu are different menu o

Pagina 30 - 4.4.2 Obtain Piece Weight

36 67XXU USB/Serial Scale Service Manual5.3 Setup Menu ParametersThis section provides more detailed descriptions of the selections found in the Setu

Pagina 31 - 4.5 Hold Function

67XXU USB/Serial Scale Service Manual 37 Table 5.2 CONFG Menu Choices and ExplanationsCONFGSubMenu1 SubMenu2 Option Default Parameter Description C

Pagina 32 - Auto HOLD

38 67XXU USB/Serial Scale Service ManualPRIM.Ut kg lb Primary unit Select the primary unit from kg or lb. The second unit is lb if kg is selected as

Pagina 33 - 4.8 Keyboard and Buzzer Test

67XXU USB/Serial Scale Service Manual 39 ZRO.PT IZSM 0-100 10 Initial zero set mechanism Range of capacity in percent± 0 - 100%FSAlso uses SAZSM and

Pagina 34

4 67XXU USB/Serial Scale Service ManualObtain Piece Weight ...

Pagina 35 - 5 Setup Mode

40 67XXU USB/Serial Scale Service Manual*The setting will be limited by the choice of REGULFUNC HOLD YES YES Enable or disable hold functionYESNO NOC

Pagina 36 - 5.3 Setup Menu Parameters

67XXU USB/Serial Scale Service Manual 41 5.3.2 USER MenuFigure 5.2 USER Menu ChartThe figure above is an illustration of the available menus with t

Pagina 37

42 67XXU USB/Serial Scale Service ManualCOM1(USB)BUD.RT 1200 9600 Select COM1 baud rate2400480096001920038400BT.FMT 8N1 8N1 Select COM1 byte format

Pagina 38

67XXU USB/Serial Scale Service Manual 43 LYOUT MULTi scp01 Set COM1 content and format (1) MULTI= the following selected item in OUT1 will be output

Pagina 39

44 67XXU USB/Serial Scale Service ManualOUT1(USB)SCL.ID YES NO Enable or disable scale ID numberPrompt is "SCALE ID"NOGROSS YES yes Enable

Pagina 40

67XXU USB/Serial Scale Service Manual 45 COM2(9 Pin)BUD.RT 1200 9600 Select COM2 baud rate2400480096001920038400BT.FMT 8N1 7e1 Select COM2 byte form

Pagina 41 - 5.3.2 USER Menu

46 67XXU USB/Serial Scale Service ManualLYOUT MULTi scp02 Set COM2 content and format (1) MULTI= the following selected item in OUT2 will be output u

Pagina 42

67XXU USB/Serial Scale Service Manual 47 OUT2(9 Pin)SCL.ID YES NO Enable or disable scale ID numberPrompt is "SCALE ID"NOGROSS YES yes Ena

Pagina 43

48 67XXU USB/Serial Scale Service ManualBEEP KEY YES YES Enable or disable beep after a key is pressed.NOCOMPR NONE IN.LMT (1) NONE = no beepL.LOW (2

Pagina 44

67XXU USB/Serial Scale Service Manual 49 HOLD AD.H.SP YES NO Enable or disable use of the high speed A/D converter after entering the HOLD modeNOHLD

Pagina 45

67XXU USB/Serial Scale Service Manual 5 1 General Information and Warnings1.1 About this ManualThis manual is divided into chapters by the chapter n

Pagina 46

50 67XXU USB/Serial Scale Service ManualOTHER NLD.RG 1-255 10 No hold range Load must return to zero in divisions to perform a print when in AUTO.1-2

Pagina 47

67XXU USB/Serial Scale Service Manual 51 5.3.3 CAL MenuFigure 5.3 CAL Menu ChartThe figure above is an illustration of the available menus with the

Pagina 48

52 67XXU USB/Serial Scale Service ManualLINE CAL.P0 Line calibration point0 Do zero point calibration. This point can't be omitted.CAL.P1 Line c

Pagina 49

67XXU USB/Serial Scale Service Manual 53 5.3.4 MISC MenuFigure 5.4 MISC Menu ChartThe figure above is an illustration of the available menus with t

Pagina 50

54 67XXU USB/Serial Scale Service Manual5.4 Exit the Setup Menu1. Power off the indicator by pressing and holding the [ON/OFF] key.2. Turn the indica

Pagina 51 - 5.3.3 CAL Menu

67XXU USB/Serial Scale Service Manual 55 6 CalibrationThe configuration/calibration button must be pushed in order to calibrate the scale. Refer to

Pagina 52

56 67XXU USB/Serial Scale Service Manual6.1.2 LinearityThe scale linearity can be calibrated with a zero reference and a full scale reference or mult

Pagina 53 - 5.3.5 TEST Menu

67XXU USB/Serial Scale Service Manual 57 6.2 Geographical Adjustment1. When GEO is selected, press the [TARE] key to enter Geographical Adjustment.2

Pagina 54 - 5.4 Exit the Setup Menu

58 67XXU USB/Serial Scale Service ManualTable 6.1 Location Code For Elevation and Latitude

Pagina 55 - 6 Calibration

67XXU USB/Serial Scale Service Manual 59 6.3 Gravity Value1. When GRAVT is selected, press the [HOLD] and [UNIT] keys to input the gravity value of

Pagina 56 - 6.1.2 Linearity

6 67XXU USB/Serial Scale Service Manual1.2 Warningsl Avoid lengthy exposure to extreme heat or cold. Your scale works best when operated at normal ro

Pagina 57 - 6.2 Geographical Adjustment

60 67XXU USB/Serial Scale Service Manual6.5 Weight Fine-tuneWith this function, the user can adjust displayed weight a little without the need of sta

Pagina 58

67XXU USB/Serial Scale Service Manual 61 7 Serial Communication7.1 Communication Settings1. Power on the indicator by pressing and holding the [ON/O

Pagina 59 - 6.4 Input or View Calibration

62 67XXU USB/Serial Scale Service Manual7.2 Com Port 1COM1 is a USB port used as a virtual RS-232 port. Communication wires come from the USB connect

Pagina 60 - 6.6 Display ADC Output Code

67XXU USB/Serial Scale Service Manual 63 Table 7.2 Bit Definition of H1H2H3 H47.6 Commands and Response7.6.1 Parameter NCI-SCP01Set the USER-COM-LY

Pagina 61 - 7 Serial Communication

64 67XXU USB/Serial Scale Service ManualCommand: W<CR> (57h 0dh), request current readingCommand: S<CR> (53h 0dh), request current status

Pagina 62 - 7.5 Transaction String

67XXU USB/Serial Scale Service Manual 65 Command: U<CR> (55h 0dh)Command: L<CR> (4ch 0dh)Command: X<CR> (58h 0dh)Command: all othe

Pagina 63 - 7.6 Commands and Response

66 67XXU USB/Serial Scale Service Manual7.6.2 Parameter ECR-SCP02Set the USER-COM-LYOUT parameter to SCP02.Command: W<CR> (57h 0dh), request cu

Pagina 64

67XXU USB/Serial Scale Service Manual 67 Command: U<CR> (55h 0dh)Command: u<CR> (75h 0dh)Command: A<CR> (41h 0dh)Command: m<CR&

Pagina 65 - Command: all others

68 67XXU USB/Serial Scale Service ManualCommand: <ENQ><CR> (05h 0dh)Command: all othersTable 7.4 Summary of Command and Response:Respons

Pagina 66 - 7.6.2 Parameter ECR-SCP02

67XXU USB/Serial Scale Service Manual 69 7.6.3 Parameter 8213-SCP03Set the USER-COM-LYOUT parameter to SCP03.Command: W (57h), request current readi

Pagina 67

67XXU USB/Serial Scale Service Manual 7 1.6 CE CertificationPending

Pagina 68

70 67XXU USB/Serial Scale Service ManualCommand: H (48h) Command: A (41h)Command: B (42h)Command: E (45h)Response Send high resolution weight dat (x1

Pagina 69 - 7.6.3 Parameter 8213-SCP03

67XXU USB/Serial Scale Service Manual 71 Command: F (46h)Command: all othersTable 7.6 Summary of Command and Response:Response Scale is taken out o

Pagina 70 - Command: E (45h)

72 67XXU USB/Serial Scale Service Manual7.6.4 Parameter MultiSet the USER-COM-LYOUT parameter to mutli.Output string frameCommand: W<CR> (57h 0

Pagina 71 - Command: F (46h)

67XXU USB/Serial Scale Service Manual 73 Example LayoutsWhen USER-OUT-xxxx is set to YES Weighing Mode:Counting Mode: SCALE ID: 123456GROSS: 123lb

Pagina 72 - 7.6.4 Parameter Multi

74 67XXU USB/Serial Scale Service Manual7.6.5 Parameter EH-SCP (PS-60)Set the USER-COM-LYOUT parameter to eh-scpThis protocol is compatible with Tole

Pagina 73 - Counting Mode:

67XXU USB/Serial Scale Service Manual 75 7.6.6 Parameter IBMSet the USER-COM-LYOUT parameter to ibm.Command: <US> W (1Fh 57h), request current

Pagina 74 - Output Status Bit Meaning

76 67XXU USB/Serial Scale Service ManualCommand: <US> H (1Fh 48h) Command: <US> A (1Fh 41h)Command: <US> B (1Fh 42h)Command: <US

Pagina 75 - 7.6.6 Parameter IBM

67XXU USB/Serial Scale Service Manual 77 Command: <US> F (1Fh 46h)Command: all othersTable 7.9 Summary of Command and Response:Response Scale

Pagina 76

78 67XXU USB/Serial Scale Service Manual8 Legal for TradeThe 67XXU Model must be configured to meet regulations mandated by local weights and measure

Pagina 77

67XXU USB/Serial Scale Service Manual 79 8.2 Audit Trail ParametersTwo separate incrementing, non-resetable audit trail parameters are used to indic

Pagina 78 - 8 Legal for Trade

8 67XXU USB/Serial Scale Service Manual2 Installation1. Unpack the scale and all components from the shipping box.2. Remove the shipping screw from t

Pagina 79 - 8.2 Audit Trail Parameters

80 67XXU USB/Serial Scale Service Manual9 TroubleshootingThis chapter gives explanations on commonly seen errors, display characters and display symb

Pagina 80 - 9 Troubleshooting

67XXU USB/Serial Scale Service Manual 81 9.2 Display SymbolsSymbol Description0---- Zero is over the setting range0_____ Zero point is below the set

Pagina 81 - 9.2 Display Symbols

82 67XXU USB/Serial Scale Service Manual9.3 Error Messages and TroubleshootingSPL.PW Sample goods weight to calculate piece weightIN.PCS Input number

Pagina 82

67XXU USB/Serial Scale Service Manual 83 PCT.ER Unit-Percentage -Weight is error, it's too small (the weight of 1%, 0.1%, or 0.01%-determined b

Pagina 84

Brecknell USA1000 Armstrong Dr.Fairmont MN 56031 Tel:507-238-8702Fax:507-238-8271Email: [email protected]://www.brecknellscales.comBreckne

Pagina 85 - Brecknell UK

67XXU USB/Serial Scale Service Manual 9 2.1 ConnectionsThere are connectors located on the side of the scale for the power supply, USB port, RS-232