67XXUUSB/Serial Scale SeriesService ManualSOFTWARE VERSION 1.20AWT35-501145 Issue AE
10 67XXU USB/Serial Scale Service ManualThe RS-232 port connector is a 9-pin female.Figure 2.3 RS-232 Connector (Com 2)2.2 Column and Remote Display
67XXU USB/Serial Scale Service Manual 11 3. Run the display cable down the inside of the column using the hole on the top side of the column and pul
12 67XXU USB/Serial Scale Service Manual5. Secure the cable to the column with the cable hold down strap that was remove in step 1.Figure 2.8 Cable
67XXU USB/Serial Scale Service Manual 13 7. Install the two screws included with the kit into the screw holes leaving them somewhat loose.Figure 2.1
14 67XXU USB/Serial Scale Service Manual10. Turn the scale upright.Figure 2.13 Column and Remote Display Installed2.3 Mounting the Display to the Sc
67XXU USB/Serial Scale Service Manual 15 2. Clip the display to the bracket.Figure 2.15 Display Attached to the Bracket3. Pull the cable all of the
16 67XXU USB/Serial Scale Service Manual6. Line up holes of the bracket with the holes on the scale base. Fasten the bracket down with the four provi
67XXU USB/Serial Scale Service Manual 17 2.4 Desktop and Wall Display MountingThe indicator is connected to the front of the 67XXU. It can be remove
18 67XXU USB/Serial Scale Service Manual2.5 MainboardFigure 2.23 shows the main board connectors for the loadcell, RS-232 port, USB port, power suppl
67XXU USB/Serial Scale Service Manual 19 Table 2.3 USB Connector (USB) Com 1Table 2.4 RS-232 Port Connector (SIO) Com 2Table 2.5 CAL Jumper SetPi
67xx Serial Scale_s_en_500145.bookBrecknell is part of Avery Weigh-Tronix. Avery Weigh-Tronix is a trademark of the Illinois Tool Works group of compa
20 67XXU USB/Serial Scale Service Manual3 Specifications3.1 Available Models3.2 Outline DimensionsFigure 3.1 67XXU Dimension DrawingFigure 3.2 67XX
67XXU USB/Serial Scale Service Manual 21 3.3 Mountingl Bracket l Remote (optional cable required)l Column (option kit required)3.4 Power Supplyl AC
22 67XXU USB/Serial Scale Service Manual3.7 Communication3.8 Analog Circuitryl 24-bit A/D converterl Conversion Speed: 10Hzl Input range: -15mV to +1
67XXU USB/Serial Scale Service Manual 23 3.12 Other Main Functionsl Programmable zero rangel Programmable pre-set tare weightl Programmable automati
24 67XXU USB/Serial Scale Service Manual4 ConfigurationTo set up the indicator, you must first enter the appropriate menu mode. The front panel keys
67XXU USB/Serial Scale Service Manual 25 4.2 KeysThe keyboard consists of four keys, some of which have multiple functions.Table 4.2 Function of th
26 67XXU USB/Serial Scale Service Manual4.3 Navigating the Weigh Mode4.3.1 Enter the Weigh ModePress the [HOLD] + [UNIT] keys for 3 seconds. Use the
67XXU USB/Serial Scale Service Manual 27 4.3.7 Output Data (print to a computer or printer)When scale is stable press the [PRINT] key.4.3.8 Display
28 67XXU USB/Serial Scale Service ManualTable 4.4 Use kg as Primary UnitTable 4.5 Use lb as Primary UnitCalibration Division ValueDisplay Division
67XXU USB/Serial Scale Service Manual 29 4.3.10 Check Weight (compare feature)To make the weight compare function available, set CONFG-FUNC-COMPAR i
67XXU USB/Serial Scale Service Manual 3 Table of ContentsChapter 1 General Information and Warnings ...
30 67XXU USB/Serial Scale Service Manual4.4.2 Obtain Piece WeightThere are two ways to obtain the piece weight: Input Piece Weight Directly1. Press t
67XXU USB/Serial Scale Service Manual 31 8. Press the [TARE] key to confirm the entry. 8a. If the calculated piece weight is less than 0.5d, the ind
32 67XXU USB/Serial Scale Service ManualPositive Peak HOLDWhen USER-HOLD-HLD.MD is set to PS.PEK, the hold mode is positive peak hold mode. When scal
67XXU USB/Serial Scale Service Manual 33 The scale will exit HOLD mode according to the setting of USER-HOLD-HLD.TM. If the time of scale variation
34 67XXU USB/Serial Scale Service Manual4.9 Serial Port 1/2 (COM1/2) Receiving TestBefore testing the receiving function of COM1 or COM2, a cable is
67XXU USB/Serial Scale Service Manual 35 5 Setup Mode The setup menu consists of five different sub-menus. Within each sub-menu are different menu o
36 67XXU USB/Serial Scale Service Manual5.3 Setup Menu ParametersThis section provides more detailed descriptions of the selections found in the Setu
67XXU USB/Serial Scale Service Manual 37 Table 5.2 CONFG Menu Choices and ExplanationsCONFGSubMenu1 SubMenu2 Option Default Parameter Description C
38 67XXU USB/Serial Scale Service ManualPRIM.Ut kg lb Primary unit Select the primary unit from kg or lb. The second unit is lb if kg is selected as
67XXU USB/Serial Scale Service Manual 39 ZRO.PT IZSM 0-100 10 Initial zero set mechanism Range of capacity in percent± 0 - 100%FSAlso uses SAZSM and
4 67XXU USB/Serial Scale Service ManualObtain Piece Weight ...
40 67XXU USB/Serial Scale Service Manual*The setting will be limited by the choice of REGULFUNC HOLD YES YES Enable or disable hold functionYESNO NOC
67XXU USB/Serial Scale Service Manual 41 5.3.2 USER MenuFigure 5.2 USER Menu ChartThe figure above is an illustration of the available menus with t
42 67XXU USB/Serial Scale Service ManualCOM1(USB)BUD.RT 1200 9600 Select COM1 baud rate2400480096001920038400BT.FMT 8N1 8N1 Select COM1 byte format
67XXU USB/Serial Scale Service Manual 43 LYOUT MULTi scp01 Set COM1 content and format (1) MULTI= the following selected item in OUT1 will be output
44 67XXU USB/Serial Scale Service ManualOUT1(USB)SCL.ID YES NO Enable or disable scale ID numberPrompt is "SCALE ID"NOGROSS YES yes Enable
67XXU USB/Serial Scale Service Manual 45 COM2(9 Pin)BUD.RT 1200 9600 Select COM2 baud rate2400480096001920038400BT.FMT 8N1 7e1 Select COM2 byte form
46 67XXU USB/Serial Scale Service ManualLYOUT MULTi scp02 Set COM2 content and format (1) MULTI= the following selected item in OUT2 will be output u
67XXU USB/Serial Scale Service Manual 47 OUT2(9 Pin)SCL.ID YES NO Enable or disable scale ID numberPrompt is "SCALE ID"NOGROSS YES yes Ena
48 67XXU USB/Serial Scale Service ManualBEEP KEY YES YES Enable or disable beep after a key is pressed.NOCOMPR NONE IN.LMT (1) NONE = no beepL.LOW (2
67XXU USB/Serial Scale Service Manual 49 HOLD AD.H.SP YES NO Enable or disable use of the high speed A/D converter after entering the HOLD modeNOHLD
67XXU USB/Serial Scale Service Manual 5 1 General Information and Warnings1.1 About this ManualThis manual is divided into chapters by the chapter n
50 67XXU USB/Serial Scale Service ManualOTHER NLD.RG 1-255 10 No hold range Load must return to zero in divisions to perform a print when in AUTO.1-2
67XXU USB/Serial Scale Service Manual 51 5.3.3 CAL MenuFigure 5.3 CAL Menu ChartThe figure above is an illustration of the available menus with the
52 67XXU USB/Serial Scale Service ManualLINE CAL.P0 Line calibration point0 Do zero point calibration. This point can't be omitted.CAL.P1 Line c
67XXU USB/Serial Scale Service Manual 53 5.3.4 MISC MenuFigure 5.4 MISC Menu ChartThe figure above is an illustration of the available menus with t
54 67XXU USB/Serial Scale Service Manual5.4 Exit the Setup Menu1. Power off the indicator by pressing and holding the [ON/OFF] key.2. Turn the indica
67XXU USB/Serial Scale Service Manual 55 6 CalibrationThe configuration/calibration button must be pushed in order to calibrate the scale. Refer to
56 67XXU USB/Serial Scale Service Manual6.1.2 LinearityThe scale linearity can be calibrated with a zero reference and a full scale reference or mult
67XXU USB/Serial Scale Service Manual 57 6.2 Geographical Adjustment1. When GEO is selected, press the [TARE] key to enter Geographical Adjustment.2
58 67XXU USB/Serial Scale Service ManualTable 6.1 Location Code For Elevation and Latitude
67XXU USB/Serial Scale Service Manual 59 6.3 Gravity Value1. When GRAVT is selected, press the [HOLD] and [UNIT] keys to input the gravity value of
6 67XXU USB/Serial Scale Service Manual1.2 Warningsl Avoid lengthy exposure to extreme heat or cold. Your scale works best when operated at normal ro
60 67XXU USB/Serial Scale Service Manual6.5 Weight Fine-tuneWith this function, the user can adjust displayed weight a little without the need of sta
67XXU USB/Serial Scale Service Manual 61 7 Serial Communication7.1 Communication Settings1. Power on the indicator by pressing and holding the [ON/O
62 67XXU USB/Serial Scale Service Manual7.2 Com Port 1COM1 is a USB port used as a virtual RS-232 port. Communication wires come from the USB connect
67XXU USB/Serial Scale Service Manual 63 Table 7.2 Bit Definition of H1H2H3 H47.6 Commands and Response7.6.1 Parameter NCI-SCP01Set the USER-COM-LY
64 67XXU USB/Serial Scale Service ManualCommand: W<CR> (57h 0dh), request current readingCommand: S<CR> (53h 0dh), request current status
67XXU USB/Serial Scale Service Manual 65 Command: U<CR> (55h 0dh)Command: L<CR> (4ch 0dh)Command: X<CR> (58h 0dh)Command: all othe
66 67XXU USB/Serial Scale Service Manual7.6.2 Parameter ECR-SCP02Set the USER-COM-LYOUT parameter to SCP02.Command: W<CR> (57h 0dh), request cu
67XXU USB/Serial Scale Service Manual 67 Command: U<CR> (55h 0dh)Command: u<CR> (75h 0dh)Command: A<CR> (41h 0dh)Command: m<CR&
68 67XXU USB/Serial Scale Service ManualCommand: <ENQ><CR> (05h 0dh)Command: all othersTable 7.4 Summary of Command and Response:Respons
67XXU USB/Serial Scale Service Manual 69 7.6.3 Parameter 8213-SCP03Set the USER-COM-LYOUT parameter to SCP03.Command: W (57h), request current readi
67XXU USB/Serial Scale Service Manual 7 1.6 CE CertificationPending
70 67XXU USB/Serial Scale Service ManualCommand: H (48h) Command: A (41h)Command: B (42h)Command: E (45h)Response Send high resolution weight dat (x1
67XXU USB/Serial Scale Service Manual 71 Command: F (46h)Command: all othersTable 7.6 Summary of Command and Response:Response Scale is taken out o
72 67XXU USB/Serial Scale Service Manual7.6.4 Parameter MultiSet the USER-COM-LYOUT parameter to mutli.Output string frameCommand: W<CR> (57h 0
67XXU USB/Serial Scale Service Manual 73 Example LayoutsWhen USER-OUT-xxxx is set to YES Weighing Mode:Counting Mode: SCALE ID: 123456GROSS: 123lb
74 67XXU USB/Serial Scale Service Manual7.6.5 Parameter EH-SCP (PS-60)Set the USER-COM-LYOUT parameter to eh-scpThis protocol is compatible with Tole
67XXU USB/Serial Scale Service Manual 75 7.6.6 Parameter IBMSet the USER-COM-LYOUT parameter to ibm.Command: <US> W (1Fh 57h), request current
76 67XXU USB/Serial Scale Service ManualCommand: <US> H (1Fh 48h) Command: <US> A (1Fh 41h)Command: <US> B (1Fh 42h)Command: <US
67XXU USB/Serial Scale Service Manual 77 Command: <US> F (1Fh 46h)Command: all othersTable 7.9 Summary of Command and Response:Response Scale
78 67XXU USB/Serial Scale Service Manual8 Legal for TradeThe 67XXU Model must be configured to meet regulations mandated by local weights and measure
67XXU USB/Serial Scale Service Manual 79 8.2 Audit Trail ParametersTwo separate incrementing, non-resetable audit trail parameters are used to indic
8 67XXU USB/Serial Scale Service Manual2 Installation1. Unpack the scale and all components from the shipping box.2. Remove the shipping screw from t
80 67XXU USB/Serial Scale Service Manual9 TroubleshootingThis chapter gives explanations on commonly seen errors, display characters and display symb
67XXU USB/Serial Scale Service Manual 81 9.2 Display SymbolsSymbol Description0---- Zero is over the setting range0_____ Zero point is below the set
82 67XXU USB/Serial Scale Service Manual9.3 Error Messages and TroubleshootingSPL.PW Sample goods weight to calculate piece weightIN.PCS Input number
67XXU USB/Serial Scale Service Manual 83 PCT.ER Unit-Percentage -Weight is error, it's too small (the weight of 1%, 0.1%, or 0.01%-determined b
Brecknell USA1000 Armstrong Dr.Fairmont MN 56031 Tel:507-238-8702Fax:507-238-8271Email: [email protected]://www.brecknellscales.comBreckne
67XXU USB/Serial Scale Service Manual 9 2.1 ConnectionsThere are connectors located on the side of the scale for the power supply, USB port, RS-232