To access the Sensor On-Off menu, press the TEST
key. The current state “ON” (orOFF”) of the photoeye
will be displayed. Pressing the ZERO key will
toggle the state ON” or OFF.” Press the TEST key
to save the setting. When the photoeye is manu-
ally turned off, you must either TEACH or turn ON
to re-enable. The display will return to the “TEACH”
prompt again.
To exit the Sensor menu, press the ZERO key until…
“DONE” is displayed.
Press TEST to return to the “FILT” message prompt
again.
From “FILT” press the ZERO key three times until…
“DONE” is displayed.
Press the TEST key to return to the “DIAG” menu
prompt.
The Calibration (CAL) Mode menu lets you cali-
brate your scale. The items in the calibration menu
are as follows:
Unit of Measure (LB, OUNCE, GR) – Selects the
operating unit of measure of your scale.
Capacity – Selects the capacity of your scale. See
available capacity/resolution selections in the
specifications section. You must choose a valid
value that is dependent on the model and capac-
ity of your unit.
Sensor On-Off
Done
(Press SW-1 to return
to normal operation.)
Step-by-Step
Instructions for
CAL Mode
**WARNING!****WARNING!**
**WARNING!****WARNING!**
**WARNING!**
Entering into this mode
will erase calibration data
already saved. Adequate
calibration weights are
required before you enter
this mode. To abort from
calibration and restore the
previous calibration data,
press SW-1 again.
17